• Clinical Chemistry Analyzer
Clinical Chemistry Analyzer

Model CA-310
  • NEW

Compact desk-top instrument with 270 tests/hour (maximum 450 tests/hour with ISE).Fully direct automatic HbA1c measurement without manual sample pretreatment.
Particularly, ideal for routine chemistry and specialized tests in small to mid-size labs.

CA-310 is a successor model of CA-270 that has been well reputed.
Compact analyzer designed for accuracy and efficiency with low reaction volume of 80 μL.

  • Healthcare

Features

Benefits provided to customers

Technologies and Systems for Precision

High-Precision liquid Dispensing System

Minimum sample volume of 1.0µl.
Liquid-level detection, pipette collision sensor and clot detection sensor ensure reliable liquid control.

Enhanced Wash System

Minimize carryover for accurate measurement.

Minimal Reaction Volume

Reaction Volume of 80µl significantly lowers reagent consumption and operational costs.

Mixing system

An effective mixing system with flexible configuration settings.

Comprehensive 13 Multi-Wavelength System

Optimal configuration for wide range of reagents.

Enhanced QC and Calibration Data Traceability

Easy access calibration history by lot based calibrator management. Both QC data and calibrator data are traceable, enhancing reliability.

Maximized operational efficiency with multifunctionality

One device, numerous test menu

From routine, special, ISE to automatic hemolysis HbA1c. Supports haemolytic, icteric, and lipemic (HIL) Detection.

Safe and Easy Sample Loading

Wide-opening cover for smooth sample addition and replacement.

Intuitive Operation and Status Display

User-friendly interface minimizes operational time for efficiency.

Streamlined Test Report management

Instant result search output by real time result and history management.

Eco-friendly power management

Automatically lowers electricity consumption with intelligent power saving mode.
Reagents remain cooled inside the unit.

Specifications

General

Type

Desktop fully automated clinical chemistry analyzer

Usage

General chemistry as photometric assay
Immunology as photometric assay (Latex reagents available)

Assay type

End point, 2 point end, rate, 2 point rate and ISE(Ion selective electrode)

Throughput

270 tests per hour (Up to 450 tests per hour with optional ISE module)
only HbA1c: 90 tests per hour

Incubation time

10 minutes after applying sample for one-reagent assay
5 minutes after applying second reagent for two-reagent assays

Sample type

Serum, plasma, urine, whole blood

Sample input system

Type : Removable tray with sample tube holder
Capacity : Up to 50 positions including 10 desinated inner positions
Sample tube size: Diameter 12 to 16 mm, height 75 to 100 mm, made of resin or glass

Sampling pipette

Type: Micro pipette dedicated for sample with liquid level sensor, clot detection and crash detection
Sample dilution: Available
Other function: HbA1c Automatic Pretreatment Measurement

Sampling pump

Sampling volume: 1.0 uL ~ 25 uL (increment by 0.1 uL)

Reagent system

Type: Removable tray with reagent bottle holder on a turntable
Capacity : 50 reagent positions on a tray (25 each for 70 ml and 20 ml bottles)
Bottle type: 70 mL bottle, 20 mL bottle
Cooling system: Cooling with peltier element (8 degrees Celcius)

Reagent pipette

Type: Micro pipette dedicated for reagent with liquid level sensor and crash detection

Reagent pump

R1 volume: 20 uL~250 uL (increment by 1 uL)
R2 volume: 5 uL~180 uL (increment by 1 uL)

Reaction system

Number of cuvettes: 72 cuvettes on reaction line
Material : Resin (Option: Glass)
light length: 5 mm
Minimum volume: 80 uL
Maximum volume: 300 uL in total
Temperature: 37±0.1 ℃

Detector

Wavelength(nm): (13 wavelengths) 340, 380, 415, 450, 478, 510, 546, 570, 600, 660, 700, 750, 800
Light source: Halogen tungsten lamp

Stirring system

Stirrer (stirrer paddle) driven by a stepping motor (in 5 speed levels)

Dimensions

870mm(W)×748mm(D)×635mm(H)

Option

ISE module, glass cuvette

Reagents for clinical chemistry

Reagents designed for automatic chemistry analyzer and optimised chemistry parameters offer the precise and reliable data to the laboratory.

  • Reagents

Clinical chemistry reagents, Immunoturbidimetric reagents

  • Calibrators

Clinical chemistry calibrator, Immunoturbidimetric calibrator

  • Quality control samples

Clinical chemistry quality control sample, Immunoturbidimetric quality control sample

  • Utility reagent

Analyzer consumables

Furuno Clinical Chemistry Analyzer

We provide extensive support to our customers based on our extensive knowledge of both reagents and instruments.

MODEL

Features

CA-310

Compact desk-top instrument with 270 test/hour (maximum 450 tests/hour with ISE).

CA-400

Mid-range desk-top instrument with 400 test/hour (maximum 560 tests/hour with ISE).

CA-800

800 tests/hour throughput capability (maximum 1200 tests/hour with ISE).

Quality and Service Information

FURUNO works daily with quality in mind, not only in the production and quality departments, but in all departments to provide users with high-quality products. FURUNO works as one company to provide safe, high-quality products and services and to achieve customer satisfaction.

Downloads

Inquiries about Healthcare Products